Scanner not sending to PACS? A DICOM connection checklist

Nine checks, in the order an integration engineer would run them, to get a modality talking to PACS again.

A new scanner arrives, or a PACS is migrated, and suddenly studies stop arriving. Most DICOM connection failures come down to a handful of settings. Work through these in order.

1. AE titles match exactly

Each DICOM device has an Application Entity (AE) title of up to 16 characters. The modality must know the PACS AE title, and the PACS must have the modality registered with its AE title. Treat AE titles as case-sensitive and check for trailing spaces.

2. IP address and port

Confirm the PACS IP address and listening port on the modality. Common ports are 104 and 11112, but many systems use their own. A static IP (or a DHCP reservation) for every modality avoids silent breakage after a power cut.

3. Run a C-ECHO (DICOM ping)

Almost every modality has a “Verify” or “Echo” button. A successful C-ECHO proves the network path, port and AE titles are right. If the echo fails, do not look further up the list until it works.

4. Firewalls and network segments

Modalities often sit on a separate VLAN. Check that the firewall allows the DICOM port in both directions where needed (storage commitment and worklist replies come back from the server).

5. Echo works but images do not arrive

Look at association negotiation. The modality proposes SOP classes (for example CT Image Storage) and transfer syntaxes (uncompressed, JPEG lossless, JPEG 2000). If the PACS does not accept that combination, the store is rejected. The PACS log usually names the refused class.

6. Modality worklist (MWL)

If technologists type patient details by hand, mistakes follow. Configure the modality to query the worklist (a DICOM C-FIND) so patient name, ID and accession number come from the RIS order. Check the worklist AE title, port and the filters (station AE title, modality type, date).

7. Patient and accession data

Studies that arrive but do not attach to the right order usually have a mismatched accession number or patient ID. Fix this at the source (the worklist), not by editing studies afterwards.

8. Storage commitment and MPPS

Storage commitment lets the modality delete local copies only when PACS confirms safe storage. MPPS reports exam start and completion back to the RIS. Both are optional but valuable; both need their own configuration.

9. Time and disk

Unsynchronised clocks confuse audit trails and worklist date filters, so point every device at the same NTP server. And check that the modality’s local disk is not full of unsent studies.